pH Analyzer MPH 66

The analyzer can measure pH in a one or two channel system. The sensors can be installed at one point or at different points up to 1000-m distance from the analyzer. The analyzer can produce an arithmetical average of both measurements. The arithmetical average can be used as an input value for the controller. Measurement can be completed with two temperature measurements. The pH value can be automatically corrected by temperature.

The analyzer contains one or two PID Controllers with width or frequency pulse output. Output can be unidirectional for solenoid valves or dosing pump control, and bi-directional for control drives with semi-continuous movement. The analyzer can be also supplied with closed loop control 0(4) - 20 mA.

The semiautomatic calibration with the automatic buffer identification and system check makes, calibration easier and prevents human error.

The analyzer is supplied in version T for indoor installation and in version V for outdoor installation.

Specifications
Measurement Range (Display) pH 0,00 to 14,00
Temperature -5 to 105°C
Sub Ranges (Analogue Output) pH 0,0 - 14,0 0,0 - 10,0
pH 2,0 - 12,0 4,0 - 14,0
2,0 - 7,0 4,0 - 9,0
7,0 - 12,0 or other
Temperature 0 - 50°C, 0 - 100°C, or other
Display 2-line alphanumeric LCD readout
16 digits per line
Output 1 up to 4 outputs 0(4) ... 20 mA
Load max. 500 Ohm
Temperature Compensation -pH automatic in range -5 to 105°C
Basic Accuracy - pH ± 0,5% of the range
Basic Accuracy - Temperature ± 0,3°C
Relay Outputs none, 2 or 4 relays (high/low)
(set points, hysteresis, delay - programmable)
voltage free contact, 250 V/50 Hz - max. 3 A
* 24 Vss - max. 0,5 A
Time Delay 0,1 až 240 minutes
Controller 2x single-parameter PID controller with
analogue or pulse output
Automatic Sensor Cleaning adjustable frequency in range of 1 minutes to 24 hours
Voltage Protection built-in lightning protection circuitry
Temperature Sensors TNiK21, TNiK22
Enclosure IP 65
Power max. 20 VA
Supply Voltage 230 V +6% to -15%, 50/60 Hz
Dimensions W 219 x H 184 x D 115 mm
Weight ca 1,5 kg
Ambient Conditions
Temperature 0 to +35°C/-20 to +35°C - version T/version V
Barometric Pressure 600 to 1060 hPa
Relative Humidity 10 ... 90%

* Valid for resistance load