Je nutné si stáhnout novou verzi flash přehrávače!

Ozone Controller MFD 88 - ozone

Up to 12 measurement channels (pH and temperature) simultaneously

Up to 8 output relays

RS 485 MODBUS, WIFI




The MFD 88 – ozone controller is a twelve channel instrument, which can measure ozone and temperature in twelve channels simultaneously. Sensors can be at different points, at a distance of up to 500 m from the controller.

Controller can be used for measurement of ozone, pH and ORP in any combination also.

All constants specific to the particular place of measurement, (e.g. calibration constants) are stored in sensors or input blocks, so it is possible change of installation place without necessity of recalibration.

Interconnection between controllers and sensor is always accomplished by ordinary non-shielded two wire cable even in case of application with automatic mechanical cleaning.

Interface of instruments comprise: Serial RS 485 Modbus (RTU + ASCII ), WIFI for easy communication with notebooks, tablets, mobiles.

The controller is provided with four analogue active current outputs. The outputs can be configured as two state also and in this mode can operate as four more relays. The total number of relays is possible increase in this way to eight.

Four programmable relay outputs can be used for signaling of exceeding of adjusted limit values or as power contacts for controlling solenoid valves, actuators or contactors in the process.

Temperature range is expanded to – 20 to 50°C

Supply voltage can be 230V, 110V 50/60Hz and 24V=

The controller is manufactured in version T - for indoor installation and in version V - for outdoor installation.

Specifications
Measurement Range - Ozone 0.01 to 10.0 mg/l (ppm)
Sub Ranges (Analogue Output) 1.0, 2.0, 5.0, 10.0 mg/l or other
Measurement Range - pH 0.0 … 14.0 pH units
Sub Ranges (Analogue Output) 0.0 – 14.0, 0.0 – 10.0, 2.0 – 12.0
4.0 – 14.0. 2.0 – 7.0, 4.0 – 9.0
7.0 – 12.0 or other
Resolution - Ozone 0,001 mg/l
Resolution - pH 0,01
Display backlighted graphic display
Outputs active 2 or 4 outputs 0(4) - 20 mA
Load max. 500 ohm
Temperature Compensation automatic
Basic Accuracy - Ozone 5% of the range at constant temperature
Relay Outputs none, 2 or 4 relays
(set points, hysteresis, delay - programmable)
voltage free contact, 250 V/50 Hz - max. 3 A
*24 Vss - max. 0,5 A
Time Delay 0,0 to 240 minutes
Enclosure IP 65
Supply Voltage 110/230 V +6% -15%, 50/60 Hz, 24V= ±20%
Power max. 20 VA
Dimensions 219x184x115 mm W x H x D
Weight cca 1,5 kg
Ambient Conditions
Temperature -20 až +50°C
input unit -20 to +70°C
Barometric Pressure 600 to 1060 hPa
Relative Humidity 10 ... 90%

* Valid for resistance load



eBRÁNA webarchitect