Je nutné si stáhnout novou verzi flash přehrávače!

Chlorine analyzers MPC 66 and MPC 66OR

The MPC 66 analyzer measures chlorine concentration either with the CSCT 43 or CSUT 43 sensors and simultaneously pH value. Chlorine concentration can also be simultaneously measured from 2 sensors (without pH measurement).

The MPC 66OR analyzer measures chlorine concentration, pH value and ORP.

The analyzer can control the dosing of bleach or chlorine gas in drinking water processing or in swimming pools. Simultaneously, it can control dosing pumps controlling pH value.

Analyzer Specifications
Measurement Range - Chlorine 0,01 to 5 mg/l
(pH 4 to 7.8 with sensor CSCT 43)
0.01 to 10.00 mg/
(pH 5.5 to 9.5 with sensor CSUT 43)
Sub Ranges (Analogue Output) 1,0-2,0-5,0-10 mg/l, or other
Measurement Range pH 0,0 to 14,0 pH units
Sub Ranges (Analogue Output) 0,0 to 14,0 0,0 to 9,0 7,0 to 12,0 or other
Measurement Range ORP ±1000 mV (MPC 66OR)
Sub Ranges (Analogue Output) ±500 mV, ±1000 mV, 0 to 500 mV, 0 to 1000 mV or other
Output Signal all values) 0/4 - 20 mA
Load Resistor max. 500 Ohm
Temperature Compensation automatic (both values) of the range 1 to 40 °C
Basic Accuracy - Chlorine ±3% of the range
Basic Accuracy - pH ±0,5% of the range
Signaling of Exceeding (Relay Outputs) none, 2 or 4 relays (high/low)
(set points, hysteresis, delay - programmable)
voltage free contact, 250 V/50 Hz - max. 3 A or
*24 Vss - max. 0,5 A
Time Delay 0,0 to 240 minutes
Recommended pH Sensor SEOJ 13
Enclosure IP 65
Supply Voltage 230 V +6% to -15%, 50/60 Hz
Power max. 20 VA
Dimensions 219x184x115 mm
Weight cca 1,5 kg
Ambient Conditions
Ambient Temperature 0 to +35°C/-20 to +35°C - version T/version V
Input Unit -20 to +70°C
Relative Humidity 10 to 90%
Barometric Pressure 600 to 1060 hPa
Supply Voltage 230V/50Hz +6%/-15%
Mains Frequency 50Hz

*Valid for resistance load