MFD 77 Multimet
Specifications
Measurement Range
|
Conductivity 0,01 uS/cm to 1000 m |
---|---|
pH 0,00 to 14,00 | |
ORP ±1000 mV | |
O2 concentration 0,1 to 30,0 mg/l (ppm) | |
O2 saturation 1 to 200% | |
Temperature T1,T2 - 5,0 to 105,0 °C | |
Temperature T3-O2 - 4,0 to 40,0 °C | |
Resolution
|
Conductivity 0,01 uS/cm to 1000 uS/cm depending on measuring range |
pH 0,01 | |
ORP 1 mV | |
O2 concentration 0,01 mg/l (ppm) | |
O2 saturation 0,1% | |
Temperature 0,1 °C | |
Display
|
illuminated graphic LCD |
Serial Output
|
RS 232 |
Analogue Output
|
0,0 to 1,0 V |
Sensor pH, ORP
|
BNC connector |
Max. Sensor Resistance
|
1.109 Ohm |
Temperature Compensation - pH
|
automatic |
Input Resistance (pH, ORP, Inputs)
|
1.1012 Ohm |
Input Current (pH, ORP, Inputs)
|
3.10-12 A |
Basic Accuracy - ORP
|
±1 mV |
Basic Accuracy -pH
|
±0,02 of pH unit |
Sensor for Temperature Measurement (T1,T2)
|
TNiK 115 |
Sensor - Oxygen Measurement
|
CTOS 43 |
Basic Accuracy - Temperature T1,T2
|
±0,2°C |
Temperature Compensation - Oxygen
|
automatic in range -1,0 to 40,0 °C |
Salinity Correction - Oxygen
|
automatic according to EN 25814 |
Barometric Pressure Compensation - Oxygen
|
automatic in range 500 to 1200 hPa |
Basic Accuracy - Oxygen
|
±1% of the range |
Additional Error at Temperature Change of Measured Medium - Oxygen
|
±2% of the range at temperature t = 20°C±15°C |
Basic Accuracy - Temperature T3
|
±0,2°C in range 0...40°C |
Sensor - Conductivity
|
CE 212, CE 213 |
Reference Temperature - Conductivity
|
25 °C |
Temperature Compensation - Conductivity
|
automatic - temperature coefficient adjustable in |
Basic Accuracy - Conductivity
|
±1% of the range ( to 10,0 mS/cm ) |
Enclosure
|
IP 00 |
Power
|
net adapter 3,6 VAC |
Dimensions
|
200x65x150 mm ( W x H x D) |
Weight
|
ca 0,5 kg |